32
CAS ELECTRICS CO.,LTD
Guatemala Guatemala 9.7844 Boston 9.8039
Hungary Budapest 9.8069 Chicago 9.8024
Indonesia Djakarta 9.7814 Dallas 9.7949
Iraq Baghdad 9.7964 Detroit 9.8039
Japan Mishima 9.7979 Los Angeles 9.7979
Korea Seoul 9.7994 New York 9.8024
Kuwait Kuwait 9.7919 Philadelphia 9.8024
Lebanon Beirut 9.7964 San Francisco 9.7994
Mauritius Port Louis 9.7859 Venezuela Caracas 9.7829
NOTE: The G-Constant is the acceleration of gravity in meters per second per second.
4.1.5 Percent Calibration (Menu Code 8150)
(Calibration MENU -> 1. Calibration -> 5. Percent Calibration)
When the case you don’t have max weight for calibration. Percent Calibration enables to use
lighter weight.
For 5kg weight, input “5”key and press “print”
Clear the tray and press “PRINT”, then “Wait4~Wait0” will display.
Put 5kg on the tray the press “PRINT” after “Wait4~Wait0” exit menu.
8100
CAL
ModE
CALIBRATION
(
2
/
3
)
4. Gravity Constant
5. Percent Calibration
6. Linearity Adjust
LoAd
0
10731
SPAN CALIBRATION
(
2
/
2
)
- Place 5.000 kg on the platter.
- Press PRINT when ready.
ULoAd
10731
10731
ZERO CALIBRATION
(1/
2
)
- Remove all weight.
- Press PRINT when ready.
8150
CAL
ModE
PERCENT
CALIBRATION
(1/
1
)
Use Weight :[15] kg
Full Capa Weight:15.000 kg

Table of Contents