CALIBRATION CHAPTER 5
5-50
6. Enter the following code into the display using the Cursor Keys: 5222
( Left / Right to select digit ; Up / Down = Increment / Decrement value ; ‘Enter’ to accept)
Calibration CW:
7. Select Calibrate CW. Press ‘ENTER’.
A warning is displayed.
Select ‘
Yes’. Press ‘Enter’ .
8. Using the Up or Down Cursor keys, enter “
1” for a single transducer .
Using the Up or Down Cursor keys, enter “
4” for the 4-in-1 transducer.
9. Press ENTER. The display reads: