71
rst me using the CEFALY DUAL Enhanced it must be fully charged. Once your
CEFALY DUAL Enhanced is fully charged, the indicator light will remain turned on.
Possible Cause:
• The sensaons experienced while using the CEFALY DUAL Enhanced are novel to
most people and can take some adjustment.
Soluon:
• To make treatment more enjoyable, we recommend you increase the intensity in
small increments of me, over the course of a few weeks. If you feel the intensity is
too strong for you, simply press the device buon once during a session to stabilize
the pulses at a tolerable level for the remainder of the session.