85

For a charge between 0 and 1000 ohms.
Constant amplitude intensity predened automacally, depending on the program.
Rectangular impulses.
Compensated constant current prevenng any galvanic direct component. Any electrode for
which the current density is above 2mA/ cm
2
requires special aenon from the operator.
Any electrode for which the current density is above 2mA/ cm
2
requires special aenon from the
operator.

When the CEFALY DUAL Enhanced is operated at an ambient temperature of 95°F the temperature
of the electrode will stay below 104°F.