PL-61044_B 12

of the UltraMIST Therapy System should assure that it is used in such an environment.
Emissions Test Compliance 
RF emissions
CISPR 11
Group 1 TheUltraMISTSystemusesRFenergyonlyforitsinternalfuncon.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The UltraMIST System is suitable for use in all establishments other
thandomesc,andmaybeusedindomescestablishmentsandthose
directly connected to the public low-voltage power supply network that
suppliesbuildingsusedfordomescpurposes,providedthefollowing
warning is heeded:
Warning: This equipment/system is intended for use by healthcare
professionals only. This equipment/system may cause radio
interferenceormaydisrupttheoperaonofnearbyequipment.It
maybenecessarytotakemigaonmeasures,suchasre-orienngor
relocangtheUltraMISTSystemorshieldingthelocaon.
Harmonic emissions
IEC 61000-3-2
Class A
Voltageuctuaons/icker
emissions
IEC 61000-3-3
Complies
Ref. IEC 60601-1-2:2007 §5.2.1.1
Emissions Test Compliance 
MaximumAcouscOutput
Power
Complies TheUltraMISTSystemrequirementfortheMaximumAcouscOutput
Power is that the measured value be less than 3W/cm2, which is a
requirementestablishedbytheWorldHealthOrganizaonforacousc
safety.Theactualoutputissignicantlylessthantheabovelimit,which
is considered to be safe.
SidewallRadiaon Complies TheUltraMISTSystemrequirementforSidewallRadiaonisthatthe
measured value be less than 100mW/cm2, which is a requirement
establishedbytheWorldHealthOrganizaonforacouscsafety.
Theactualoutputissignicantlylessthantheabovelimit,whichis
considered to be safe.
Ref. TR-75143 on le at Celleraon

