GENERATOR
1 Generator
2 Power Switch (Back of Generator)
3 Generator Power Cord
4 Peristaltic Pump
5 Tubing Guides
6 Adjustable Hook
7 Indicator Light
8 Treatment Wand Connector
9 Treatment Wand Cradle
10 USB Port/Cover
TREATMENT WAND
11 Treatment Wand
12 Display Screen
13 Keypad
14 Treatment Wand Cable with Connector
15 Tubing Channel
16 Treatment ON/OFF Button (Trigger)
17 Treatment Wand Tip
APPLICATOR
18 Applicator
19 Tubing-Cable Clip
20 Tubing Clamp
21 Infusion Tubing
22 Tubing Spike
23 Tubing Spike Protective Cover
PLEASE READ INSTRUCTIONS BEFORE USE.
UltraMIST
System
ULTRAMIST SYSTEM INSTRUCTIONS FOR USE
Other Materials Required
• 0.9% Normal Saline 100cc - 1000cc bag
• Germicidal wipes for cleaning/disinfecting
• Absorbent pads
Caution: R
x
Only. This medical device is to be operated by a healthcare professional in accordance with the instructions for use.
Generator
Treatment Wand
Applicator
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
PL-61044_B 1

Questions and Answers

  • B
    Bruce PriceAug 16, 2025
    What to do if my Celleration Ultramist system will not turn on?
    • A
      Amy SandovalAug 16, 2025
      If your Celleration Medical Equipment system isn't turning on, make sure the power cord is properly plugged into both the generator outlet on the back of the instrument and the power outlet. Also, verify that the power outlet is delivering power. If the issue persists, contact Celleration Customer Service, as the fuse on the system may be missing or blown.
  • T
    Thomas HowellAug 19, 2025
    What to do if no saline is coming out of the Celleration Ultramist applicator?
    • K
      Kristina WoodsAug 19, 2025
      If no saline or insufficient saline is coming out of the Celleration Medical Equipment applicator, ensure that the 'wound ready' screen is visible on the display. Check that the treatment on/off button has been pressed to begin the treatment. Make sure the saline pump door is closed and the tubing is properly loaded in the saline pump using the tubing guides. Also, verify that the tubing is not clamped, kinked, or knotted. Finally, ensure the treatment wand tip is correctly aligned in the center of the applicator and that the tubing spike is fully mated with the saline bag.
  • C
    Christian PhelpsAug 22, 2025
    What to do if my Celleration Ultramist Medical Equipment treatment wand has no power?
    • V
      Veronica LewisAug 22, 2025
      If there is no power to the treatment wand of your Celleration Medical Equipment (and no screen display), first ensure that the system is turned on using the power switch located on the back of the system. Also, verify that the treatment wand cable is securely inserted into the connector on the generator.
  • A
    Ashley EscobarAug 25, 2025
    What does it mean if my Celleration Ultramist Medical Equipment says 'Overheat'?
    • O
      Omar LeeAug 26, 2025
      The 'Overheat' alarm on your Celleration Medical Equipment indicates that the ultrasound is being turned off as a preventive measure because the treatment wand temperature has exceeded the allowed limit. This does not necessarily indicate a malfunction. The system will begin a cool-down process, counting down the minutes required to cool the system. Do not turn off the system during this cool-down period, as it could require a replacement of the applicator. Once the cool-down is complete, the 'paused treatment' screen will appear, and you can resume the treatment. Note that 20 minutes is the expected maximum treatment time after periods of system inactivity.
  • J
    jocelynpatrickAug 28, 2025
    What to do if saline is leaking around the Celleration Medical Equipment applicator?
    • G
      Gary MooreAug 28, 2025
      If you notice saline leaking around the applicator of your Celleration Medical Equipment, ensure that the treatment wand tip is properly aligned in the center of the applicator. Also, if the treatment wand position during treatment caused saline to accumulate inside the applicator, turn the applicator tip downward to allow excess saline to drain and then continue treatment.
  • P
    popeemilyAug 31, 2025
    What does it mean if I see an error message on my Celleration Medical Equipment display screen?
    • C
      catherine45Aug 31, 2025
      If you see an error message on the display screen of your Celleration Medical Equipment, refer to the common errors table to determine the appropriate action. If the error continues, note the contents of the error message and call Celleration Customer Service.
  • G
    gloverkellySep 3, 2025
    What to do if all keys on the Celleration Ultramist keypad are not functioning?
    • E
      Elizabeth RobertsSep 3, 2025
      If all the keys on your Celleration Medical Equipment keypad are not functioning, shut down the system by turning off the power and then restarting it. If the error continues, call Celleration Customer Service, as this may indicate a system software error.
  • E
    estradadavidSep 6, 2025
    What to do if my Celleration Medical Equipment applicator is damaged?
    • B
      Barbara SmithSep 6, 2025
      If your Celleration Medical Equipment applicator is damaged, replace it with a new applicator. The damage might have occurred during shipment.
  • J
    Jessica Lewis DDSSep 9, 2025
    What to do if a single key on the Celleration Ultramist Medical Equipment keypad is not functioning?
    • M
      melindaandersonSep 9, 2025
      If a single key on the keypad of your Celleration Medical Equipment is not functioning, call Celleration Customer Service, as this may indicate a mechanical failure.
  • R
    Roger SmithSep 12, 2025
    What to do if the sterile Celleration Medical Equipment applicator package is open or damaged?
    • J
      jenniferbraunSep 12, 2025
      If the sterile applicator package for your Celleration Medical Equipment is open or damaged, replace it with a new applicator, as the damage might have occurred during shipment.

Summary

Please Read Instructions Before Use

Description and Indications for Use

Contraindications and Potential Complications

Preparing the Patient for Treatment

Cleaning the UltraMIST System

Preparing the UltraMIST System

Treating the Patient with the UltraMIST System

UltraMIST System Shutdown Procedure

USB Port Information

Common Errors and Indicator Lights

Common Errors

Lists common error messages and their causes on the display screen.

Indicator Light Status

Explains the meaning of the indicator lights on the generator.

Troubleshooting

System Will Not Turn On

Troubleshooting steps for when the system fails to power on.

No Power to Treatment Wand

Solutions for issues where the treatment wand shows no display or power.

Keypad Not Functioning

Addresses problems with individual or all keypad keys not responding.

Overheat Condition

Information and solutions for the system's overheat alarm.

Display Error Messages

General guidance for interpreting and resolving display error messages.

Saline Flow Issues

Troubleshooting steps for issues with saline flow from the applicator.

Saline Leaking Around Applicator

Solutions for problems causing saline leaks around the applicator.

Damaged Applicator

Guidance on handling situations where the applicator is damaged.

Damaged Applicator Package

Instructions for dealing with damaged sterile packaging of applicators.

Warnings and Precautions

General Warnings

Important safety warnings regarding device operation and handling.

Usage Precautions

Precautions to ensure treatment effectiveness and proper device use.

Product Specifications and Technical Service

Product Specifications

Technical specifications of the UltraMIST system, including safety and performance data.

Technical Service Contact

Contact information for technical support and customer service.

System Functions and Symbols

Frequently Used Functions

Overview of commonly used controls and their functions.

Explanation of Symbols

Definitions and explanations of symbols used throughout the manual.

Manufacturers Declarations

Electronic Emissions Declaration

Declaration regarding the system's electronic emissions and compliance with standards.

Radiation Emissions Declaration

Declaration regarding the system's radiation emissions and safety.

Federal Communications Commission (FCC) User Information

Celleration Ultramist Specifications

General IconGeneral
Frequency40 kHz
PortabilityPortable
ApplicationWound care
TechnologyLow-frequency ultrasound
Power SourceAC adapter