Preparing the UltraMIST System continued on next page
PREPARING THE ULTRAMIST™ SYSTEM
Withthepowerswitchinthe‘o’posion,plugthe
supplied generator power cord into the back of the
generator and into an electrical outlet.
1
Ensurethetreatmentwandcableisaachedandfully
seated to the front of the generator by aligning the red
mark on the cable with the red mark on the connector.
2
Turn the power switch located on the back of the
generatortothe‘on’posion.
3
Aertheself-testiscompleted,usethekeypadonthe
treatment wand to select the number of wounds to be
treated. Press toconnue.
4
Use the keypad to enter the size of each wound to be
treated per the screen prompts.
5
The total saline requirement for the combined treatment
area is displayed (approximately 100mL for every 25
cm
2
). Select the appropriate saline bag size. Press to
connue.
6
The display screen will ask for the applicator to be
loaded. Remove applicator from packaging and press
the applicator onto the treatment wand. Ensure the
applicatorisfullyseatedpriortousewiththemetalp
visible in the center.
7
Note: Aer the set up is complete, DO NOT remove the applicator
unl the treatment has been fully concluded. This is a single use
product that once removed will need to be replaced with a new
applicator.
Up
Down
Back
Select
(Forward)
-- Treatment TimeWound Number --
-- Wound Size
PL-61044_B 3
20-29
ULTRAMIST™ SYSTEM INSTRUCTIONS FOR USE