PROBLEM CAUSES SOLUTIONS
NO SALINE OR INSUFFICIENT
SALINE IS COMING OUT OF
APPLICATOR
Wound ready screen is not visible on
display screen
screen is seen on display
for treatment to begin
Ensure system power is on and press treatment
Saline pump door is open Close pump door
Tubing is not properly loaded in the
saline pump
Load tubing into saline pump using tubing
guides and close pump door
Tubing is clamped
Unkink or remove knot from tubing
aligned in center of applicator
applicator opening and fully seated on the
applicator housing
Tubing spike not fully mated with
saline bag
Ensure spike and saline bag are fully mated
SALINE LEAKING AROUND
APPLICATOR
aligned in center of applicator
applicator opening
treatment caused saline to
accumulate inside applicator
DAMAGED APPLICATOR Damaged in shipment Replace with new applicator
OPEN OR DAMAGED STERILE
APPLICATOR PACKAGE
Damaged in shipment Replace with new applicator
TROUBLESHOOTING
PL-61044_B 9
ULTRAMIST™ SYSTEM INSTRUCTIONS FOR USE