COMMON ERRORS (noted on display screen)
# ERROR CAUSE SOLUTIONS
008 OVERHEAT Treatment wand temperature has
exceeded the allowed limit
the system
20 minutes is the expected maximum treatment
to count down the minutes required to cool
the system. Once this completes, the “paused
treatment” screen will appear and the treatment
can now be completed.
DO NOT TURN OFF SYSTEM during cool down (this
will require a replacement of the applicator)
200 INVALID
REPLACE APPLICATOR
Applicator has been previously used
and should be replaced
Replace applicator
201 REPOSITION
APPLICATOR
treatment wand
Adjust current applicator properly on treatment
wand
202 MAX USE
REPLACE APPLICATOR
Applicator has been used for the
203 NOT READING
APPLICATOR
Applicator is not being sensed by
treatment wand
Properly align the applicator on the treatment
wand