Celleration Ultramist - Common Errors and Indicator Lights; Common Errors; Indicator Light Status

Celleration Ultramist
13 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMON ERRORS (noted on display screen)
# ERROR CAUSE SOLUTIONS
008 OVERHEAT Treatment wand temperature has
exceeded the allowed limit



the system
20 minutes is the expected maximum treatment


to count down the minutes required to cool
the system. Once this completes, the “paused
treatment” screen will appear and the treatment
can now be completed.
DO NOT TURN OFF SYSTEM during cool down (this
will require a replacement of the applicator)
200 INVALID
REPLACE APPLICATOR
Applicator has been previously used
and should be replaced
Replace applicator
201 REPOSITION
APPLICATOR

treatment wand
Adjust current applicator properly on treatment
wand
202 MAX USE
REPLACE APPLICATOR
Applicator has been used for the


203 NOT READING
APPLICATOR
Applicator is not being sensed by
treatment wand
Properly align the applicator on the treatment
wand


INDICATOR LIGHT (located on front of generator)
RED A
GREEN 
PL-61044_B 7
ULTRAMIST SYSTEM INSTRUCTIONS FOR USE