TREATING THE PATIENT WITH THE ULTRAMIST™ SYSTEM
The treatment wand should be oriented so that it is perpendicular to the
wound with the leading edge of the applicator approximately 0.5 to 1.5 cm
(0.2-0.6 inches) from the wound.
Depressandreleasethetreatmenton/obuon(trigger)onthetreatment
wandhandle.Thiswilliniateboththesalineowandultrasounddelivery.
Thetreatmentmerwillstarttocountdownonthedisplay.
1
The UltraMIST System will begin to deliver the ultrasound through a
connuousmist.Movethetreatmentwandvercallyacrossthewound
usingmulplepasses.Followthisbymovingthetreatmentwand
horizontallyacrossthewoundusingmulplepasses.Thetreatmentwand
should be moved across the wound using slow, even strokes.
Audible (gurgling) and surface visual bubbling may occur. This is normal
when moving back and forth through the recommended treatment range.
For best results, the treatment wand should be perpendicular to the wound
andparalleltotheoorduringtreatment.
Note: It is important to maintain a 0.5 to 1.5 cm (0.2 – 0.6 inch) distance from
the leading edge of the applicator to the treatment area at all mes to maximize
therapeuc eect.
3
Connuetreatmentunl3audibletonessoundindicangtheendof
treatmentme.Theultrasoundandsalinewillautomacallystop,ensuring
accuratetreatmentmeforthewoundsizeselected.
If there is more than one wound to treat, the next wound screen will
appear.Pressthetreatmenton/obuontostarttreangthenextwound.
Note: if at anyme during the treatment it is determined addional me will be
required, use the up/down buons to add me in increments of 1 minute. Press
to connue.
5
Note: You may pause treatment by depressing and
releasing the treatment on/o buon. The treatment
mer will be paused unl restarted.
0.5 - 1.5 cm
13:00
Wound 2
Acve
3:00
Treating the Patient continued on next page
PL-61044_B 5
2
4
0.5 - 1.5 cm
ULTRAMIST™ SYSTEM INSTRUCTIONS FOR USE