TREATING THE PATIENT WITH THE ULTRAMIST SYSTEM
The treatment wand should be oriented so that it is perpendicular to the
wound with the leading edge of the applicator approximately 0.5 to 1.5 cm
(0.2-0.6 inches) from the wound.
Depressandreleasethetreatmenton/obuon(trigger)onthetreatment
wandhandle.Thiswilliniateboththesalineowandultrasounddelivery.
Thetreatmentmerwillstarttocountdownonthedisplay.
1
The UltraMIST System will begin to deliver the ultrasound through a
connuousmist.Movethetreatmentwandvercallyacrossthewound
usingmulplepasses.Followthisbymovingthetreatmentwand
horizontallyacrossthewoundusingmulplepasses.Thetreatmentwand
should be moved across the wound using slow, even strokes.
Audible (gurgling) and surface visual bubbling may occur. This is normal
when moving back and forth through the recommended treatment range.
For best results, the treatment wand should be perpendicular to the wound
andparalleltotheoorduringtreatment.
Note: It is important to maintain a 0.5 to 1.5 cm (0.2 – 0.6 inch) distance from
the leading edge of the applicator to the treatment area at all mes to maximize
therapeuc eect.
3
Connuetreatmentunl3audibletonessoundindicangtheendof
treatmentme.Theultrasoundandsalinewillautomacallystop,ensuring
accuratetreatmentmeforthewoundsizeselected.
If there is more than one wound to treat, the next wound screen will
appear.Pressthetreatmenton/obuontostarttreangthenextwound.
Note: if at anyme during the treatment it is determined addional me will be
required, use the up/down buons to add me in increments of 1 minute. Press
to connue.
5
Note: You may pause treatment by depressing and
releasing the treatment on/o buon. The treatment
mer will be paused unl restarted.
0.5 - 1.5 cm

13:00
Wound 2
Acve
3:00
Treating the Patient continued on next page
PL-61044_B 5
2
4
0.5 - 1.5 cm
ULTRAMIST SYSTEM INSTRUCTIONS FOR USE