BIO X6 User Manual v. 1.0 140
Auto calibration is not
available.
Auto calibration is
not supported with
the following
printheads:
EMD Printhead
Thermoplastic
Printhead
Auto calibration is
not supported with
the following
toolheads:
Photocuring
HD Camera
Reach customer support if auto
calibration is not available with the
following printheads:
Pneumatic Printhead
Temperature-controlled Printhead
Syringe Pump Printhead
Check the sensor settings; make sure
it is not being triggered.
Print stops when
connected to external
air supply.
The air supply is
interrupted.
Check the connection and make sure
it is connected correctly without
leakage.
If compressor is being used, ensure
the compressor works as intended
and supplies sufficient pressure.
Printbed does not
reach set
temperature.
The ambient air is
too hot.
Turn off the BIO X6 and let it rest for
20 minutes.
Open the door.
Decrease the room temperature or
increase the airflow to reduce the
temperature of the ambient.
Turn on the chamber fan to faster
reduce the temperature of the BIO X6.
Place the BIO X6 in a cool room.
Temperature-
controlled Printhead
The ambient air is
too hot.
Decrease the room temperature or
increase the airflow to reduce the
temperature of the ambient air. Turn