BIO X6 User Manual v. 1.0 28
7. When you have calibrated all 6 printheads, tap Done.
The calibration trimming is now complete. The only time you need to repeat it is when you
upgrade or downgrade the firmware (HeartOS).
If you are working with a specific nozzle/needle gauge, and the autocalibration trimming has
been performed using another nozzle/needle gauge, repeat the procedure with the new
nozzle/needle to improve the accuracy of the autocalibration.