BIO X6 User Manual v. 1.0 85
Note: The Automatic calibration protocol is not compatible with the HD Camera toolhead or
the Photocuring toolhead. If one of these toolheads is connected to the BIO X6 when
performing Automatic calibration, you can choose either to calibrate them manually or to link
their calibration to one of the other toolheads.
ABL probes the bed at three points to enable bed-levelling compensation. It uses a matrix to
compensate for tilt in the bed or print surface. It is recommended to perform ABL when
printing large models or when using well plates as the printing surface.
The surface probe only needs to be calibrated once, the printer will then remember this
calibration and apply it for each printing. However, the ABL must be enabled before each
printing since it compensates not only the bed but also the print surface tilt.
Figure 58 Calibration options Droplet