3. Program Control Commands
3.3 Instrument Subsystem Command
253
Example: :CORRection:FLATness:SELect “test.plt” selects the compensation file
text.plt in the path ”SgData/user/Pltc/”
Reset state: None
Key path: [Power]>[User Calibration Compensation] >[Select Calibration
Compensation File]
Description: for setting only.
[:SOURce[1]|2]:CORRection:FLATness[:FILE]:STORe <FileName>
Function description: This instruction is used to save the current compensation
file data in another specified file. The parameter is a string type, which
indicates the saved file name. There is no need to specify a path name
in the parameter. The default path "SgData/user/Pltc/" parameter needs
to include the file extension ".plt". If the saved file name already exists,
the original file will be overwritten. This command is for setting only.
Setting format: [:SOURce[1]|2]:CORRection:FLATness[:FILE]:STORe “filename”
Parameter description:
<FileName> Character string type parameter, name of the compensation file to be
saved
Example: :CORRection:FLATness:STORe “test.plt” saves the current
compensation data to the file
”SgData/user/Pltc/”.
Reset state: None
Key path: [Power]>[User Calibration Compensation] >[Edit Calibration
Compensation File] >[Save As]
Description: for setting only.
[:SOURce[1]|2]:CORRection:FLATness:FREQuency <Val>{,{Val}}
Function description: This instruction is used to set the value of the frequency
column in the compensation file. This instruction supports setting the
frequency values of multiple lines of the frequency column. When
modifying the frequency values of multiple lines, you need to input
multiple frequency values in turn, separated by commas. If the number
of input parameters a is greater than the number of rows b in the current
compensation file, A-B sweep points are inserted in the list, the power
compensation values of newly inserted frequency points are the default
values, and the frequency values of all points are the input frequency
values. The maximum number of parameters entered is 65,535, and the
minimum number is 1. Before calling this command, you need to
confirm that a compensation file has been selected. Please refer to
"[:SOURce[1]|2]:CORRection:FLATness[:FILE]:SELect" for the
command to select the compensation file. Otherwise, a prompt of "user