Level Commands
BPFIL? Bypass Status Query
Input: BPFIL?
Returned: 0 or 1
Comments: Used to query the current hot gas bypass status where 0 = not
bypassing and 1 = bypassing.
FILAS? Fill Time Alarm Status Query
Input: FILAS?
Returned: 0 or 1
Comments: Used to query the Fill Time Alarm Status where 0 = off and 1 = on.
FILL Set Fill Status
Input: FILL 0 or FILL 1
Returned: No response
Comments: Used to set the current Fill Status where 0 = Stop Fill and 1 = Start
Fill
FILL? Fill Status Query
Input: FILL?
Returned: 0 or 1
Comments: Used to query the current Fill Status where 0 = not filling and 1 =
filling.
FILLM? Auto Fill Status Query
Input: FILLM?
Returned: 0 or 1
Comments: Used to query the Auto Fill Status where 0 = disabled and 1 =
enabled.
FILT Set Max Fill Time
Input: FILT XXX
Returned: No response
Comments: Used to set the maximum fill time in minutes from 30 to 240
minutes.
FILT? Max Fill Time Query
Input: FILT?
Returned: XXX
Comments: Used to query the maximum fill time. Value returned in minutes.
FILTIM? Fill Duration Query
Input: FILTIM?
Returned: XXX
Comments: Used to query the duration of the current fill. Value returned in
seconds.
HFIL Set High Level Setpoint
Input: HFIL XXX.X
Returned: No response
Comments: Used to set the High Level Setpoint. Input value uses the current
level units, either inches or millimeters. There must be at least a 0.5 inch interval
between each level setting and alarm.