Guidance and manufacturer’s declaration - Electromagnetic immunity - Enclosure port
Phenomenon
Basic EMC standard
or test method
Immunity test levels
Professional
healthcare facility
environment
Home healthcare
environment
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
± 8kV contact
± 2 kV, ±4kV, ±8 kV, ±15 kV air
Radiated RF EM elds IEC 61000-4-3
a
)
10 V/m
b
) 80MHz - 2.7
GHz 80% AM at 1kHz
Proximity elds
from RF wireless
communications
equipment
IEC 61000-4-3 COMPLIANT
NOTE: Further information about distances to
be maintained between portable and mobile
RF communications equipment (transmitters)
and the Chicco nebulizers Super Soft Family
00009814000000 can be requested from
Artsana S.p.A. using the contact information
provided in this manual. However, it is advisable to
keep the electromechanical aerosol equipment
at an adequate distance of, at least, 0.5 m from
mobile phones or other RF communications
transmitters to minimise possible interference.
RATED power frequency
magnetic elds.
IEC 61000-4-8 30 A/m
c
)
50 Hz or 60 Hz
a
) The equipment is suitable for use in Home Health Environments and Professional Health Care
Environments limited to patient rooms and respiratory treatment facilities in hospital or clinics.
The more restrictive IMMUNITY acceptance limits have been considered and applied.
b
) Before modulation is applied.
c
) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT or ME
SYSTEM and sources of power frequency magnetic elds.
Guidance and manufacturer’s declaration - Electromagnetic immunity - Input a.c. port
Phenomenon
Basic EMC standard
or test method
Immunity test levels
Professional
healthcare facility
environment
Home healthcare
environment
Electrical fast
transients / Bursts.
IEC 61000-4-4 ± 2 kV
100 kHz repetition frequency
Surges Line-to-line IEC 61000-4-5 ± 0.5 kV, ±1kV
Surges Line-to-
ground
IEC 61000-4-5 Class II equipment or supplied with Class II AC/
DC adapter. Test not applicable