Nitrogen Generator NiGen GC Nigen MICRO
5
2.2.
Notes on the use of nitrogen
Nitrogen is not a toxic gas, but when the percentage in the air exceeds specific
values there is a risk of asphyxia.
Therefore NEVER DIRECTLY INHALE the gas produced and avoid working in the
vicinity of a flow of nitrogen.
However, given the low quantities produced, it is sufficient that the flow of ni-
trogen occurs in a normally ventilated environment to avoid the risk of accumu-
lation.
m
If the generator and the user utility are located in different rooms, the
following minimum ventilation must be provided to prevent increasing
or decreasing the oxygen content in the air:
2.3.
Safety devices
MAXIMUM PRESSURE: the control system constantly monitors the pressure.
In case of anomalous pressure, a safety valve (which operates independently of
the control system) prevents it exceeding 10.5 bar (152 psi).
MAXIMUM TEMPERATURE (
HC version only
): the purifier is fitted with an
internal catalyser, the temperature of which is constantly maintained at 420°C
To guarantee safety, a completely independent device is fitted which shuts off
power supply to the heating element when temperature values exceed 480°C
2.4.
Technical assistance
The CLAIND technical assistance can be contacted as follows:
Phone ++39 0344 56603
Fax ++39 0344 56627
Email: service@claind.it
Website: filling the form on the website www.claind.it at the “Service” section
Model
Minimum generator
room ventilation
Minimum point-of-use
room ventilation
NiGen GC
NiGen GC HC
9 Nm3/h
318 SCFH
16 Nm3/h
565 SCFH
NiGen MICRO
3 Nm3/h
106 SCFH
5 Nm3/h
177 SCFH