Nitrogen Generators Nitro35 Nitro70
23
6.8.2.
Setting of “Minimum nitrogen pressure” alarm
MINIMUM OUTLET PRESSURE
xx.x bar
This value represents the nitrogen pressure in the internal reservoir of the ge-
nerator under which the
INSUFFICIENT NITROGEN PRESSURE.
alarm is
activated.
To access from main menu, press the Function key . This value can be incre-
ased or decreased pressing the Function key and selecting the arrow key
up or down .
To go back on main page, press the Function key . for some seconds.
e
The default value is 6.0 bar.
6.8.3.
Outlet nitrogen pressure regulation
The regulator is positioned on the front of the generator (see par. 3.3.1).
Please rotate the knob to regulate the pressure to the desired value (clockwise
to increase the pressure) displaying it on the gauge installed on the left.
e
The default value is 0.0 bar (outlet closed).