24
Symptom Probable Cause Corrective action
Display does
not work when
the keys witch
is turned on
1.Bad connection of power
cable
2.Emergency shut
-off knob is
engaged
3.Fuse is short
-circuited
4.LCD is malfunctioned.
1.Check
if the power cable is
Properly plugged into the
wall outlet and main body's
socket.
2.Turn off the system, turn
the emergency shut
-
off knob
clockwise and restart the
system.
3.4. Call your local
representative agency
Cartridge
dose
not working
when the
loaded
handpiece.
1. No display frequency
value on main screen
2. No “READY Setting
3.
Handpiece is defective
4.
Cartridge is defective
1
,2
Check connecting of
handpiece cable and
cartridge reload.
3,4 . Call your local
representative agency
No
emission of
Ultrasound
energy
1. Main body defective
2. Cartridge defective
3. Shot switch defective
1,2 ,3
Call your local
representative agency.
Troubleshooting