CM probe English 2019/02 DK
Preparation of the check calibration:
Please prepare the check calibration thoroughly:
Use the set of accessories consisting of a small plastic
box, a special liquid, filter paper and a special foil.
Clean the probe thoroughly with
alcohol and make sure the probe head
has dried completely before continuing.
Put the filter paper into the small
plastic box. Moisten the filter paper with the special liquid.
Surplus liquid can be poured away carefully. Take care that your hands are oil free
and the plastic box is clean.
For the first measurement (low values), cut out a marked piece of the special foil
and smoothly cover the moistened filter paper with it (make sure the foil has no
“wrinkles” and there are no air bubbles underneath).
Take care that the probe does not get in
contact with any liquid. Place the probe vertically
onto the foil and do not move it during the
measurement.
In the MPA-software:
Press “Start” in the software.
5 measurements are performed one after each other.