What to do if CoaguSense Medical Equipment shows 'NO SAMPLE DETECTED'?
- JJohn ButlerAug 20, 2025
If the CoaguSense Medical Equipment displays 'NO SAMPLE DETECTED', it means either no sample or not enough sample was applied to the strip within 2 minutes after the “Apply Sample” message appeared. Repeat the entire procedure with a new strip, including a fingerstick on a different finger. Make sure to apply the sample within 2 minutes of seeing the “Apply Sample” message and ensure the pipette tip or transfer tube touches the sample well before dispensing the sample.