QC Test Lipid
Lipid
Lipid
1
Material needed:
2 Lipid discs
3
Power on instrument.
Lipid
QC Info Disc
2
1 QC solution level 1
1 QC solution level 2
1 Lipid QC Info disc.
4
Select Control Test
and then select QC
Test.
9
Close instrument lid.
The QC information
will be saved into the
system.
10
Target value and target
value range are displayed.
Validate by selecting
.
11
Select level of QC test
(Level 1 or 2).
12
Open disc pouch and
remove Lipid disc.
Lipid
QC Info Disc
6
For new QC lot: Take
Lipid QC Info disc from
the QC solution kit.
5
For a new QC lot select
New QC Lot. Previously
tested QC lots can be
selected via lot button in
display.
7
Open the lid by
selecting Open on the
screen and follow the
on screen instructions.
8
Place QC Info disc
on spindle within the
white circle.