13
Open disc completely
and turn disc on its
backside (blank side
facing up).
14
Carefully apply a drop
of QC solution (level
as defined in
11
) to the
suction point.
Lipid
Lipid
16
Press the hinge cover down
firmly to close disc. Once
closed correctly the hinge
cover cannot be opened
again.
21
Open lid by selecting
Open on the screen,
remove first Lipid disc,
insert second Lipid disc,
and close lid.
22
Verify that the QC solution
lot number matches the lot
number displayed on the
screen. Select
to begin
the measurement.
24
Finish test by by
selecting on the
screen and dispose of
disc appropriately.
Lip id
23
When measurement is
complete the results of
both levels (1+2) are
displayed.
* If alarm is enabled.
15
Observe that the drop of
QC solution is completely
absorbed by the disc. The
area marked in blue has to
be filled completely with the
clear solution.
17
Open lid by selecting
Open on the screen,
remove QC Info disc,
insert Lipid disc, and
close lid.
18
Verify that the QC solution
lot number matches the lot
number displayed on the
screen. Select
to begin the
measurement.
Lip id
19
When measurement is
complete, select
to
proceed with the QC
test of the other level.
20
Open disc pouch and
remove second Lipid
disc. Repeat procedure
shown in steps 14-16.