Quick Reference Guide
Routine Testing
Power on the h 232
Press the “Patient Test” Button
Enter or scan the patient ID
Insert the test strip
The test strip will warm up
automatically
Apply the sample within 5
minutes. Then press the check
button.*
Draw exactly 150 μL sample
using a Roche CARDIAC Pipette
in between the 2 blue arrows
Use heparinized venous whole
blood only
Gently swirl the tube for
10 times
* Hold the pipette vertically over the sample
application area. Slowly apply the sample to
avoid spillage and losing of samples.
The Roche CARDIAC Pipette tip should not touch
the sample application area of the test strip.
Video Guide
Technical Support:
+852 5729 6447 (Max Tsang)
+ 852 5729 3919 (Jeffrey Chan)
Roche Diagnostics (Hong Kong) Limited

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)