What to do if scaling tip vibrates but no spray from Coltene BioSonic S1?
- WwilliamsjonathanAug 12, 2025
If the scaling tip of your Coltene Media Converter vibrates but there's no spray when you press the foot switch, it could be due to a minimal water control knob setting. Try increasing the water supply using the water control knob. Alternatively, the water system might be blocked, so clean the water line and filter.