49
5. SPECIFICATION
C9551 C9553 C9555 C9557
PSI Range 0 to ±2.031 0 to ±5.076 0 to ±30.46 0 to ±101.05
Resolution 0.001 0.001 0.001 0.01
mBar Range 0 to ±140 0 to ±350 0 to ±2100 0 to ±7000
Resolution 0.1 0.1 0.1 1
Bar Range 0 to ±2.100 0 to ±7.000
Resolution 0.001 0.001
inH
2
O Range 0 to ±56.2 0 to ±140.5 0 to ±843.1 0 to ±2810
Resolution 0.01 0.01 0.1 0.1
inHg Range 0 to ±4.134 0 to ±10.34 0 to ±62.01 0 to ±206.7
Resolution 0.001 0.001 0.01 0.01
mmHg Range 0 to ±105.0 0 to ±262.5 0 to ±1575 0 to ±5250
Resolution 0.01 0.01 0.1 1
Torr Range 0 to ±105.0 0 to ±262.5 0 to ±1575 0 to ±5250
Resolution 0.01 0.01 0.1 1
Pa Range 0 to ±14000
Resolution 1
KPa Range 0 to 14.00 0 to ±35.00 0 to ±210.0 0 to ±700.0
Resolution 0.001 0.01 0.1 0.1
mmH
2
O Range 0 to ±1428 0 to ±3569
Resolution 0.1 1
cmH
2
O Range 0 to ±142.8 0 to ±2141 0 to ±7138
Resolution 0.01 0.1 1
Kgcm
–2
Range 0 to ±143 0 to ±0.357 0 to ±2.141 0 to ±7.138
Resolution 0.001 0.001 0.001 0.001
CALIBRATED ±0.2% of full scale (FS) at +23°C
ACCURACY
REPEATABILITY ±0.1% of full scale (FS) at +23°C
MEAN TEMPERATURE <0.1% per °C
COEFFICIENT
OF READING
OPERATING 0 to 50°C, 10 to 90% RH
RANGE non condensing
ENVIRONMENTAL IP67
RATING