Troubleshooting
ECG fault
Fault Phenomenon
Possible Reason Inspection and Repair
No ECG waveform can be detected, no
heart rate value, and sometimes there
is an ECG technical alarm
Incorrect lead type setting
Check the various settings of the ECG to confirm that the
settings are correct
Poor contact between electrode pads and human body Replace pads, or clean skin, use couplant
Electrode pads and lead wires are damaged Replacement accessories
Internal module connection failure Check that the internal connections are correct
Parameter module failure Replacing the parameter module
ECG has clutter, waveform distortion,
and inaccurate heart rate
External devices are interfering with the monitor
Check the external environment to eliminate
electromagnetic interference
Inappropriate filter settings Set to appropriate filter settings
Parameter module failure Replacing the parameter module