Maintenance
Maintenance schedule
3 monthsPer day
If stored for a long time, charge the
battery to full
Clean the surface
Device works well
Accessory works well
2 years
Check:
Safety check
ECG check
NIBP leakage test, NIBP verification
IBP calibration
EtCO2 calibration
Lithium battery’s performance, replace it if needed