IE-HGS-H5(6)4XX-05
15
Regulator address setting
Regulator address for preset communication protocol
RS485 network regulator address
Communication speed
Communication speed for preset communication protocol
communication speed setting
Temperature unit choice
It is possible to measure temperature at °C or °F. There is shown preset unit. It is possible to change
it after „SET“ key pressing.
preset to °C
preset to °F
CO
2
measurement mode selection
This item provides to choice if measured value is displayed and read in “SLOW” or “FAST”
mode. “FAST” mode provides last measured value without averaging; this means that no software
filter is used for calculating the output value. “SLOW” mode provides value averaged from last 11
“FAST” measured values.
averaged value „SLOW“ mode
current value without averaging „FAST“ mode