Compliance Products USA, Inc
CAL-CDT-002-PRO CDT-240 V2 Rev 0
Page 1 of 12
SIGN OFF
DATE
ORIGINATOR:
QUALITY ASSURANCE:
DOCUMENT EFFECTIVE DATE:
1.0 TITLE
Calibration procedure for CDT-240 V2.
2.0 PURPOSE
To describe the CDT-240 V2 calibration procedure.
3.0 DEFINITIONS
DMM - Digital Multimeter
EUT Equipment under test
4.0 PRECAUTIONS
4.1 The CDT-240 V2 tester can generate high voltage. Be very careful to follow all directions
in the procedure.
5.0 MATERIALS/EQUIPMENT NEEDED
5.1 Calibrated DMM (Qty 2) DMM1 for voltage RMS below 50V and the DMM2 for
voltages RMS 50V or greater.
5.2 Calibrated Oscilloscope (with 10x or 100x probe)
5.3 DC power supply 60V
5.4 BNC cable, 1:1 ratio
5.5 Screwdrivers
5.6 CDT-Test Voltage Plug
6.0 PROCEDURE
6.1 Once a technician has been trained on this procedure by an appropriate person, follow the
directions listed in Appendix 1.
6.2 An appropriate person shall be any technician that has performed the calibration of a
CDT-240 V2 tester and can explain the procedure and safety concerns.
7.0 APPENDICES
7.1 Appendix 1 CDT-240 V2 Calibration Procedure.
8.0 RELATED ITEMS
8.1 CAL-CDT-002-LOG CDT-240 V2

Other manuals for Compliance West CDT-240

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Compliance West CDT-240 and is the answer not in the manual?

Compliance West CDT-240 Specifications

General IconGeneral
Input Voltage100-240 VAC
Input Frequency50/60 Hz
Impedance50 Ohms
DisplayLCD
Display TypeTFT LCD
Operating Temperature0°C to 50°C
Storage Temperature-20°C to +70°C