12
3.0
MAINTENANCE
Regular and proper maintenance of your
Power-
Pro
attachments are the best way to protect your
investment. It is essential that you have your
powered surgical instruments serviced as sched-
uled so as to retain their optimum performance
and reliability, which will reward you with safer,
less problematic product performance over time.
The recommended maintenance interval for ser-
vice of your
PowerPro
attachments is every 24
months.
Proper care and handling of the instruments on a
day-to-day basis, which includes proper day-to-
day operation, cleaning, and sterilization, are
extremely important to ensure safe and efficient
operation.
Your authorized Hall Surgical service depart-
ment is the most knowledgeable about the
Pow-
erPro
attachments and will provide competent
and efficient service. Service at Hall Surgical at
the recommended service intervals is mandatory
to keep your product warranties in effect. Any
services and/or repairs done by any unautho-
rized repair facility may result in reduced per-
formance of the attachments or attachment
failure and is not recommended. See
“4.3 Lin-
vatec and Hall
®
Surgical Instrument War-
ranty” on page 19
for more information on
product warranties.
3.1 Cleaning and Sterilizing
3.1.1 Cleaning Precautions
1. Follow universal precautions for protective
apparel when handling and cleaning con-
taminated instruments.
2. Do not lubricate any attachments.
3. Never immerse attachments.
4. Never clean attachments with
bleach, chlorine-based detergents, liquid or
chemical disinfectants, or any products con-
taining sodium hydroxide (such as
INSTRU-KLENZ or Buell Cleaner). They
will degrade the anodized aluminum coat-
ing.
5. Never clean equipment in an ultrasonic
cleaner or a combination washer/sanitizer.
6. Saw blades and bits are single-use
only. Do not resterilize. Dispose of
properly after use.