17
display will show the difference in %, the middle sub-display will
show the offset and the right sub-display the preset reference.
g) MEM (= Memory)
Up to 10 measurements (reference number 0 to 9) can be stored and
recalled.
Proceed as follows:
1. Push function button until MEM flashes in the display, afterwards
press the key SET/R once. MEM is fixed and the reference number
flashes. Make your measurements and push the SET/R key once to
store the present measurement value in the first free memory loca-
tion = reference no. 0.
2. Press UP/DOWN button to go to the next free memory location
(reference no. between 0 and 9).
Make again your measurement and push the key SET/R once. The
selected memory now is addressed. If a memory no. which has been
used before is selcted the previous value is updated and stored with
the new measurement value. The memorized value appears on the
middle sub-display. To exit this function mode turn the function
switch (observe the safety rules!) or press the buttons FUNC or SET/R.
Hint!
If you exit the function mode because you pushed the SET/R button too
often, this function mode is left but the memory is not erased and can
be “recalled“ with the following function mode.
h) RCL (= Recall)
This function enables you to get the memorized reading back on the
LCD.
Follow these steps:
Push FUNC key until RCL flashes. Push SET/R once. RCL is fixed but the
reference number flashes. Press UP or DOWN buttons to address the
desired number where the measurement has been memorized. Press
SET/R button to get the memorized value back on the display. The
recalled value (ref. no. 0) will appear on the left sub-display, 2nd
16
return from this function mode, press Set/Reset key twice (RESET) and
you will return to Auto Range.
e) Auto Hold A-H
In this function the meter will automatically record the minimum and
the maximum display reading, MIN and MAX values. These values are
currently updated = refreshed. The MIN-value is shown on the left
sub-display, the MAX-value on the right sub-display. On the middle
sub-display the average value = AVG is shown. This value is currently
“refreshed“. Press FUNC key until A-H flashes in this function mode.
Press once SET/R button. The symbol A-H is fixed, now the R-H (Range
Hold) symbol flashes. Now push the UP and DOWN key to select the
range for MIN, MAX and AVG measurement (e.g. for car voltage con-
trol the 40 V-DC range = decimal point in the middle). If you fixed the
range, press SET/R once again, the R-H sysmbol stops flashing. Both
symbols are visible in the top line of the display. You can start measu-
rement.
To exit this function mode press SET/R again or press FUNC or turn the
rotary function switch (observe the safety rules!), afterwards return to
normal settings.
f) REL (= Relative)
The relative mode enables the operator to compare the reference
value with a subsequent reading.
Proceed as follows:
• Set the function key at REL mode and press once the button SET/R
• Store the polarity by pressing the UP and DOWN keys, afterwards
press SET/R button.
• Press UP and DOWN buttons again to set the reference value you
desire and the decimal point (measuring range = R-H = Range-
Hold). Press UP / Down buttons and SET/R by turns to move to the
next digit.
• For final setting of the reference value press SET/R button once.
The meter will now display the difference between the stored refe-
rence value and the subsequent readings on the sub-displays, while
the present (true) measurement is on the main-display. The left sub-