1.1 Safety Instructions
The following safety instructions appear in this manual. Please read them carefully
before operating the sterilizer.
1.1.1 Aborting a Cycle
WARNING: RISK OF NON-
STERILE DEVICE
Please note that if a cycle is aborted prior to successful termination, it must
be considered incomplete and the load must be reprocessed (unwrapped
goods or liquids) or repacked and reprocessed (wrapped goods).
The MKII Controller Operator’s Guide
7
To abort a cycle in progress, press the red POWER key on the keypad assembly.
This will shut off power to the unit. You will then need to open the green–handled
manual exhaust valve (labeled Manual Steam Exhaust Valve) visible by removing
the access panel below the chamber door. Wait for the pressure inside the chamber
to reach atmosphere before carefully opening the door.
1.1.2 Starting a Cycle
WARNING: BURN HAZARD
The operator MUST wear protective clothing including face shield, thermal gloves,
and proper lab attire, when attempting to load and unload the chamber materials.
Steam released from the sterilizer chamber can cause serious burns. Stand away
while opening the door.
If water leaks from the front of the sterilizer, DO NOT open the door. Contact
service personnel.
!
!