Enteral Feeding Pump
Directions For Use
NORMAL OPERATION
Plug the pump into AC wall outlet.
BATTERY OPERATION
Unplugging the pump from the AC line will place the pump on battery power. To
recharge the battery, plug the pump into an AC wall outlet. Pump will automati-
cally begin charging.
PROGRAMMING-These general instructions are for inserting the Kangaroo
924 Feeding Sets into the Kangaroo™ 924 Feeding Pump.
1. Close roller clamp on pump set.
2. Fill feeding bag or spike container prior to setting up pump.
3. In order to achieve proper accuracy, the fluid line in the feeding bag must be
28 inches above the top of the feeding pump when initiating the feeding cycle.
4. Slowly open control clamp and fill entire line with fluid, leaving a few drops
of fluid in the drip chamber.
NOTE: Avoid filling the drip chamber more than half full.
5. Close roller clamp.
6. Load set per arrow diagram on pump. Do not overstretch tubing.
7. Press and wait while pump performs a system check.
8. Check VOL CLR light or press VOL then VOL CLR to clear volume.
9. Set feed rate using one of the two following options:
• Pressing the increase or decrease button, or
Turning the knob on the left side of the pump
Setting the Volume to be Delivered “VTBD” Rate (optional feature)
• Press the VTBD button.
• Set “VTBD” by using knob on side of pump, or by pressing the
increase or decrease button.
• Only use the “VTBD” option if you want the pump to stop and alarm once a
set amount of formula is delivered.
10 . Open roller clamp.
11. Press START/HOLD .
TO CHANGE RATE:
1. Press .
2. Set new rate by pressing the or keys on the front panel.
3. Press .
See pump operating manual for detailed instructions.
Kangaroo
924
ON
t
t
t
t
t
t
START
HOLD
START
HOLD
START
HOLD

Questions and Answers

  • B
    burnsangelaAug 18, 2025
    How to fix Flow Error Alarm (FLo Err) on Covidien Medical Equipment?
    • A
      aaron19Aug 18, 2025
      If the Covidien Medical Equipment displays a 'Flow Error Alarm (FLo Err)', it could be due to several reasons: * The feeding container might be empty, in which case you should refill it. * The feeding tube or feeding set tubing could be occluded; locate and correct the occlusion. * The drip chamber might not be properly placed in the pump; ensure it's correctly locked in the drip chamber guide. * The drip chamber walls may be coated with feeding formula; try manipulating the chamber to remove the formula, or replace the set. * The sensors in the upper drip chamber guide might be blocked or dirty; clean them with a cotton swab dampened with warm soapy water. * The drip chamber might be too full; ensure the formula isn't preventing the sensors from operating correctly.
  • A
    Adrian JuarezAug 20, 2025
    What to do if Covidien Medical Equipment shows 'Hold Error Alarm (HLd Err)'?
    • Z
      Zoe PattonAug 20, 2025
      If the Covidien Medical Equipment triggers a 'Hold Error Alarm (HLd Err)', it means the pump has been left in hold mode for more than 5 minutes. To resolve this, press START to silence the alarm, then press START again to stop the HOLD flashing, and finally, press START again to restart the pump.
  • K
    kristen61Aug 23, 2025
    What causes a 'System Error Alarm (SYS Err)' on Covidien Medical Equipment?
    • J
      Jessica RodriguezAug 23, 2025
      If the Covidien Medical Equipment displays a 'System Error Alarm (SYS Err)', it could be due to the rotor not turning while the pump is running, or the rotor turning but not being detected by the rotor sensors. In either of these cases, return the pump for technical service.
  • J
    Jessica NicholsAug 26, 2025
    What to do if Covidien Medical Equipment displays 'No Set (LoAd SEt)'?
    • Z
      Zachary FitzgeraldAug 26, 2025
      If the Covidien Medical Equipment shows a 'No Set (LoAd SEt)' error, it might be because the pump set has been improperly placed onto the pump. Ensure that only a Kangaroo™ 924 Pump Set is used. If 'dots' appear with the message, it indicates there's too much ambient light; move the pump away from bright light.
  • M
    melissaingramAug 28, 2025
    Why does my Covidien Medical Equipment have a 'Low Battery Alarm (Lo bAt)'?
    • C
      Cassandra RuizAug 28, 2025
      If the Covidien Medical Equipment is showing a 'Low Battery Alarm (Lo bAt)', it means the battery has been run down below the point of maintaining the pump's accuracy. To resolve this, plug the pump's power cord into a wall outlet.
  • A
    aaroncastilloAug 30, 2025
    Why is there No Audible Alarm on my Covidien Medical Equipment?
    • K
      Katie CarrilloAug 30, 2025
      If your Covidien Medical Equipment has no audible alarm, it could be due to a fully discharged battery. In this case, charge the battery for at least 12 hours. If charging the battery doesn't solve the issue, the problem may be alarm circuitry failure and the pump should be returned for technical service.
  • M
    Mercedes LloydSep 5, 2025
    What to do if my Covidien Kangaroo 924 displays a No Set error?
    • S
      Stacey AllenSep 5, 2025
      If your Covidien Water Pump shows a 'No Set (LoAd SEt)' error, it could be due to: * The pump set being improperly placed onto the pump. Ensure that only a Kangaroo 924 Pump Set is used. * Too much ambient light. Move the pump away from bright light if 'dots' appear with the message.
  • C
    Courtney BrownSep 7, 2025
    Why is my Covidien Water Pump showing a Low Battery Alarm?
    • T
      Timothy DavidSep 7, 2025
      If your Covidien Water Pump is displaying a 'Low Battery Alarm (Lo bAt)', it means the battery has run down to a level that affects the pump's accuracy. To resolve this, plug the pump's power cord into a wall outlet to recharge the battery.
  • A
    Andrea BoydSep 10, 2025
    What to do if my Covidien Kangaroo 924 Water Pump has a Hold Error Alarm?
    • J
      Jennifer StuartSep 10, 2025
      If you are getting a 'Hold Error Alarm (HLd Err)' on your Covidien Water Pump, it means the pump has been left in hold mode for longer than 5 minutes. To fix this, press START to silence the alarm, then press START again to stop the HOLD flashing. Finally, press START again to restart the pump.
  • S
    Sharon ChavezSep 12, 2025
    What does System Error Alarm on my Covidien Water Pump mean?
    • C
      Carlos DoyleSep 12, 2025
      If your Covidien Water Pump displays a 'System Error Alarm (SYS Err)', it indicates that either the rotor is not turning while the pump is running, or the rotor is turning but not being detected by the rotor sensors. In this case, the suggested solution is to return the pump for technical service.

Summary

Kangaroo 924 Enteral Feeding Pump Operation

Power and Battery Management

Connect to AC power for operation and battery recharging.

Feeding Set Installation and Priming

Procedure for loading the feeding set, priming the fluid line, and ensuring accuracy.

Setting Feed Rate and Volume

Configure pump rate and optional volume to be delivered (VTBD).

Starting and Stopping the Pump

Initiate feeding, pause, and restart the pump operation.

Kangaroo 924 Enteral Feeding Pump Troubleshooting

Flow Error Alarm (FLo Err)

Addresses issues like empty container, occluded tubing, or improper drip chamber placement.

Low Battery Alarm (Lo bAt)

Troubleshooting steps for low battery indication, including recharging.

Hold Error Alarm (HLd Err)

Resolves issues when the pump is left in hold mode for extended periods.

No Set (LoAd SEt) Alarm

Corrects problems with improper pump set placement onto the pump.

System Error Alarm (SYS Err)

Guidance for rotor not turning or not detected by sensors.

No Audible Alarm

Troubleshoots fully discharged battery or alarm circuitry failure.