Safety
3-2
3.2 Warnings and cautions
The warnings and cautions for the Cryomech cryocooler are listed here by section. The same warnings
and cautions appear in each respective section.
3.2.1 Section 5: Specifications
Section 5.2 Technical specifications
CAUTION
Operation of the cryocooler in any situation that does not meet the specifications in this section will void
the warranty. If you plan to operate the system outside any of the specified conditions, contact Cryomech,
Inc.
CAUTION
The use of DI (De-Ionized) water is prohibited. DI water causes damage to the compressor package and
will not be covered under warranty.
3.2.2 Section 6: Installation
Section 6.1 Cold head installation
CAUTION
Failure to follow these installation guidelines will void the warranty.
CAUTION
Special care must be taken when handling the cold head. The warranty will not cover dented or bent
tubes.
CAUTION
Operating the cold head in a magnetic field of greater than 500 gauss will void the warranty.
Section 6.1.2 Initial preparation of the cold head
CAUTION
Heating any part of the cold head above 325 K (125 ºF, 52 ºC) will cause damage and void the warranty.
CAUTION
With the cold head oriented vertically, the maximum allowable load cannot exceed the value specified in
Section 5. Excessive loading will damage the cold head. This type of damage is not covered by the
warranty. If your application requires a load heavier than the weight specified in Section 5, prior written
authorization from Cryomech, Inc. must be granted.