Cygnus 3 Operating Manual
25
Membrane Temperatures
The standard polyurethane membranes are suitable for measuring
surfaces with temperatures up to 75°C.
For measuring on higher surface temperatures Teflon membranes
are available, they are suitable for temperatures up to 150°C. See
accessories on page 121.
When measuring hot surfaces reduce the probe contact time
to a minimum to avoid damaging the probe face and crystal.
Probe Selection & Specifications
This is the standard probe
Suitable for measurement on
surfaces are relatively rough
narrower beam makes this
Probe ideal for measuring
small-bore tubing, thin section
plate and other areas where
Ideal for thin sections without
Lower frequency probes offer better penetration on heavy
corrosion/coatings.
Probe Frequency Identification
The frequency of Cygnus probes is indicated by colour;
1
To measure thicknesses on tall thin cylinders or columns the height-width ratio should be no
less than 1.0:0.6 (Height:Width) otherwise side reflections prevent measurement.