Only operate the Gauge with Deep-Coat turned On when
measuring through protective coatings thicker than 6 mm.
Make sure Deep-Coat is turned Off when measuring metal with
coatings less than 6mm.
To turn the Deep-Coat function on or off:
1.
Press the MENU key to display
the MAIN MENU
2.
Scroll down to GAUGE SETUP
then press OK
3.
Scroll down to DEEPCOAT
4.
Press the EDIT key to change
the DEEPCOAT setting:
5.
The gauge will automatically go
back to the measuring screen.