5-3. Simulation Calibration Mode (Without Test Weight)Mode 2.
- This calibration Method will be useful to make calibration more than 10ton capacity setting.
- Guaranteed resolution will be 1/5,000 and if you need higher resolution, please make
calibration with Test weight.
Step 1. Enter to the SET-CALmode
Press
key during 5 sec display.
Or turn on the Power + with pressing
display
Press
key display.
Press
key to start “Simulation Calibration Mode”.
Remarks : Go to next step with save
key / Back to previous step key Step 2.
Step 2. Digit (Division)setting ( display)
Press
key and stop the optimal Digit value.
(Whenever pressing
key, digit value will be changed, like 01-02-05-10-20-50)
Press
key and save change and move to next step.
d _ 1
CAL_ _2
SET-CAL
TEST
SET-CAL
15 / 49