Dartin Bilateral Reduction/Night Kit
Intended Use:
For collecting drainage uids from two
catheters into one bag. Provide simple
exchange from Day Bag to Night Bag.
Instructions For Use:
1 Attach end of 20” silicone tubing from Part A to
drainage catheter 1 (leur lock)
2 Remove tubing from top of DS600 (not provided)
and reattach to open arm of Y connector on Part A.
Attach other end of tubing (leur lock) to drainage
catheter 2.
3 Attach Part B (friction t) to top of DS600.
4 Connect leur lock from top of part B to
female leur lock on Part A.
To Switch to Night Bag:
1 Clamp off tubing at bottom of Part A. Using proper
sterile technique, undo connection between Part A and Part B.
2 Attach tubing (leur lock) from DNLL2000 Night Bag to Part A.
3 Release clamp on Part A.
Sterile if package is unopned and undamaged. For single use only.
The reuse of this single-use device can lead to patient infection and/or device malfunction.
Caution Federal Law (USA) resticts this device to sale by or on the order of a physician.
Dartin Medical Systems Canada
To Reorder: Call 1-800-585-2941 Fax: (905) 403-8631
EN
Anti-Reflux Inlet
-600-
-500-
-400-
-300-
-200-
-100-
Dartin Medical
800-585-2941
CAUTION:
CLOSE/TURN EMPTY PORT
CLOCKWISE BEFORE USE
Approx. ml
Dartin Select 600 ml
EOSTERILE
2
Drainage Catheter 2 Drainage
Catheter 1
20” Silicone Tubing
from DS600
Y Connector
Clamp
Part A
Female Leur Lock
Connector
Part B
DS600 with
Part B attached
20” Silicone
Tubing from
Part A
Question and Answer IconNeed help?

Do you have a question about the Dartin Medical Systems Dartin Select 600 ml and is the answer not in the manual?

Dartin Medical Systems Dartin Select 600 ml Specifications

General IconGeneral
BrandDartin Medical Systems
ModelDartin Select 600 ml
CategoryMedical Equipment
LanguageEnglish

Summary

Instructions For Use

Attach Tubing to Catheter 1

Attach the 20-inch silicone tubing from Part A to drainage catheter 1 (leur lock).

Reconnect Tubing to Y Connector and Catheter 2

Reconnect tubing from DS600 to Y connector on Part A and attach to catheter 2.

Attach Part B to DS600 Drainage Bag

Securely attach Part B to the top of the DS600 drainage bag using a friction fit.

Connect Part B to Female Leur Lock on Part A

Connect the leur lock from the top of Part B to the female leur lock on Part A.

To Switch to Night Bag

Clamp Tubing and Disconnect

Clamp the tubing at the bottom of Part A and disconnect Part A from Part B.

Connect Night Bag Tubing

Attach the tubing from the DNLL2000 Night Bag to Part A using the leur lock.

Release Clamp on Part A

Release the clamp on Part A to allow fluid flow.