10 | www.deepelec.com | NanoVNA-F Handheld Vector Network Analyzer Incomplete Manual v1.0
Using RF cables to connect Port 0 and port 1, view S21 curve errors should not exceed 0.1dB.
If calibration data abnormalities are found, they should be re-calibrated.
If calibration is available, the CAL status will be displayed. Otherwise, it is hidden.
C* is in the state where an unsaved calibration is applied (it disappears from memory when the power is
turned off). C0 to C4 Indicates that the saved calibration values are applied to one of the save locations.
Saved or Unsaved will change when calibration data is properly saved.
The letters below C indicate that the following error terms have been applied.
D: Directivity, R: Reflection Tracking, S: Source Match, T: Transmission Tracking and X: Isolation
The instrument has been fully frequency-calibrated(50kHz~1GHz) before shipment and Store the data
in parameter 0, and the boot will load the status 0 data directly.
If you change the measurement conditions, such as adding a adapter and modifying the frequency range,
you may need to recalibrate and store the parameters in 0-4. All parameters can be called up by the “recall”
function.