  • K
    kennethduffySep 3, 2025
    Why is the galvanization temperature too low in DeguDent SOLARIS Medical Equipment?
    • B
      breanna08Sep 4, 2025
      If the galvanization temperature is significantly too low during the process with your DeguDent Medical Equipment, it could be due to using too much liquid, a defective heater, or a defective heater controller. Use the suggested liquid volume (set in user menu).
  • C
    Caitlyn MartinSep 18, 2025
    How to resolve 'EEPROM read error' on DeguDent SOLARIS?
    • D
      Debbie OconnellSep 18, 2025
      If the DeguDent Medical Equipment displays an EEPROM read error, and the device number in EEPROM is not legible, enter the serial number in EEPROM using the ewi command. Alternatively, try restarting the unit.
  • K
    Kimberly SantiagoSep 16, 2025
    How to fix a defective temperature sensor on DeguDent Medical Equipment?
    • K
      Kiara RiveraSep 16, 2025
      If the DeguDent Medical Equipment indicates a defective temperature sensor, this is likely due to a hardware problem. Replacing the temperature sensor is the suggested solution.
  • C
    Cassandra McmahonSep 14, 2025
    Why is the menu very dark on my DeguDent Medical Equipment?
    • J
      jacobsstevenSep 14, 2025
      If the menu on your DeguDent Medical Equipment is very dark, the screen saver may be active. Use a control (button, control knob) to exit the screen saver.
  • D
    David EvansSep 12, 2025
    Why does the selection move slowly on my DeguDent SOLARIS?
    • Y
      ymorrisSep 12, 2025
      If the selection moves slowly or discontinuously on your DeguDent Medical Equipment, it could be due to an incorrect encoder type being selected. Use the eww command to adapt the encoder type in EEPROM to the actual hardware (optical encoder, mechanical encoder).
  • R
    Richard LawrenceSep 12, 2025
    What to do if DeguDent SOLARIS Medical Equipment doesn't recognize the reference electrode?
    • D
      David SotoSep 12, 2025
      If the reference electrode is not recognized by your DeguDent Medical Equipment, make sure that the reference electrode was placed and check its contacts. Also, ensure that the anode and galvanizing liquid are present.
  • J
    joshua68Sep 9, 2025
    Why doesn't galvanization start above 65°C in DeguDent SOLARIS?
    • M
      Michael GarciaSep 9, 2025
      If galvanization does not start on your DeguDent Medical Equipment despite a temperature display significantly above 65°C, the galvanization temperature estimate might be incorrect. Check the galvanizing temperature EEPROM value using the rhg command; adjust that value using the shg command if required.
  • M
    Matthew HarperSep 7, 2025
    What causes high galvanization temperature in DeguDent Medical Equipment?
    • R
      rtaylorSep 7, 2025
      If the galvanization temperature is significantly too high during the process with your DeguDent Medical Equipment, the heater controller may be defective. Replace the heater controller.
  • A
    Ashley MillerSep 5, 2025
    What to do if objects hit the beaker rim in DeguDent SOLARIS?
    • B
      Brad HollandSep 5, 2025
      If objects hit the rim of the beaker as the slide is lowered in your DeguDent Medical Equipment, it could be due to using the wrong beaker (small rather than large) or incorrect positioning (outside rather than inside). Use the beaker shown on the loading menu.
  • W
    William SanchezAug 20, 2025
    What to do if the DeguDent SOLARIS Medical Equipment shows 'Motor malfunction. Move head up'?
    • D
      darren89Aug 21, 2025
      If the DeguDent Medical Equipment displays a 'Motor malfunction. Move head up' error, it could be due to an obstacle. Try to remove any obstacles that may be hindering the movement. Alternatively, attempt to restart the movement.

DeguDent SOLARIS Specifications

General IconGeneral
CategoryMedical Equipment
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Beam LimitationRectangular collimation
Source-to-Skin Distance20 cm
X-ray Tube Voltage70 kV
Power Supply50/60 Hz
ManufacturerDeguDent