4
TO SET THE SCALE
Set the scale to # 1 for most paper and paper products, # 2 for obtaining relative moisture indications on
hygroscopic materials for which no established calibration is available, or # 3 for baled scrap paper.
To change the scale, press and hold the scale button (#4). The meter will display the current
scale for one second then scroll forward through the scales.
Release the button to stop at your desired scale.
Changing the scale will automatically reset the set-point value to the default setting for that particular
scale. Default settings are as follows:
Scale #1 - 7%
Scale #2 - 50
Scale #3 - 19 %
TO CHANGE THE SET-POINT
To change the set-point value press the set-point button (#3). The meter will display the
current set-point value for the scale you have chosen tor one second.
To scroll forward to a higher value for that scale hold the set-point button (#3) while the
current set-point is displayed and scroll to the set-point value desired.
To scroll backward through the set-point values, press and release the set- point button (#3).
Within one second, press and hold the scale button (#4). Continue to hold the scale button (#4)
and the set-point wilt decrease.
When scrolling in either direction, release the button to stop at your desired set-point.
A buzzer sounds if the meter reads a %MC higher than the set-point.
You can change the value between 5.0 and 18.0 for Scale #1, 2-99 for Scale #2, and between 6.0 and
39.0 for Scale #3.
INFORMATION ABOUT YOUR READINGS
Readings below the nominal range of each scale will be displayed as a negative number. Readings above
the nominal range will be displayed with a blinking
number. All under-range and over-range readings should be disregarded. They are not added to the
accumulated readings or used in calculation of the average or highest reading.
The meter can accumulate up to 1 00 readings. After all 1 00 readings are stored, it will not add new
readings until the memory has been cleared. lt will also continue to display the average of all 100
readings as a reminder that the memory is full.
To add a reading to the sum of all the previously stored readings, release the read button (#l)
within 2 seconds.
When taking and storing readings for a specific material, be sure to clear the meter before moving on to
the next scale if you do not want to group all of the readings together.

Other manuals for Delmhorst P-2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Delmhorst P-2000 and is the answer not in the manual?

Delmhorst P-2000 Specifications

General IconGeneral
BrandDelmhorst
ModelP-2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals