INTRODUCTION
The DELTA OHM DO 9721 quantum photo-radiometer and thermometer data logger has been
designed for measuring illuminance, radiance, luminance and temperature.
The instrument has two inputs, A and B, and automatically recognises the probes connected,
whether they be illuminance, radiance, luminance or temperature probes, and can display the diffe-
rence between the two inputs.
As the probes are interchangeable, it is possible to choose the most suitable combination for all
applications without having to recalibrate the instrument. The DO 9721 is able to take illuminance
measurements in lux and in fcd (foot-candle), irradiance measurements in W/m
2
, in µW/cm
2
e in
µmol/m
2
s, luminance measurements in cd/m
2
and temperature measurements in °C or °F.
The instrument’s Data Logger function enables it to store up to 30,000 readings. The sampling
period is variable from 1 second to 12 hours.
The data acquired may later be transferred to a Personal Computer or a printer by means of the
opto-insulated serial line RS232C. For each value stored the date and time of acquisition are indi-
cated; each acquisition block is ended with a report which provides the maximum, minimum and
mean values.
With the Serial Output function it is possible to obtain the instantaneous values measured by the
instrument at the output of the serial line RS232C, in order to send them to a printer or a computer.
Other functions such as Hold (which blocks the display), Rel (for taking relative measurements),
Record (for storing the maximum, minimum and mean values) and Q (integration in time of the
measurements with alarm threshold) further enrich the instrument’s performance.
Thanks to its versatility and to its storage capacity, the instrument is suitable for a wide variety of
applications, both in the field and in the laboratory.
79
ENGLISH