MiniMouse II
TM
Mini
Centrifuge
Operations Manual
Introduction:
Thank you for purchasing the MiniMouse II™ Mini
Centrifuge. Please read this manual thoroughly
prior to operating the instrument.
The MiniMouse II™ Mini Centrifuge is designed for
low speed applications and quick spin downs of
microcentrifuge tubes (0.2ml - 2.0ml) and for
eight position PCR strips.
Specifications:
Speed: 6000 rpm / 2000xg
Capacity: 8 x1.5/2.0ml
32 x 0.2ml or 4 x PCR strips
Dimensions: 4.5 (W) x 5.9 (D) x 4.5 (H) in.
11.4 (W) x 15 (D) x 11.4(H) cm
Electrical: 120V to 240V, 50-60 Hz
Max. Radius: 4.9cm
Installation:
Place the MiniMouse II™ on a flat and level
surface nearby an electrical outlet. It should not
be near a heat source or in direct sunlight.
Rotor Exchange:
The MiniMouse II™ includes a second rotor that is
designed for holding PCR tubes (0.2ml) and PCR
strips (8 position). The rotor is conveniently
stored in a compartment on the bottom of the
centrifuge. To access this rotor, turn the
machine upside down and remove it from the
compartment.
- Removing the Rotor from the
motor shaft:
The new Clik-n-Lok rotor attachment
system allows for instant exchange of
rotors. To remove a rotor, simply pull it
upwards, off the shaft. (NO TOOL IS
REQUIRED.)
- Attaching the Rotor to the motor
shaft: Place the rotor on the motor
shaft and press down until the rotor
comes to a stop. The rotor is now
secured. (NOT TOOL IS REQUIRED)
Operation:
- Attach the power cord and place the power
switch into the on “-” position.
- Open the lid and load your samples into the
rotor. Always ensure a balanced load
(Please see the following section, Loading
the Rotor, for details.)
- Close the lid, the rotor quickly accelerates
to the maximum speed of 6200 rpm.
- When the desired run time is completed,
push on the lid button to open the lid, the
rotor quickly brakes to a stop and the
samples can be retrieved.
NEVER Attempt: to remove samples until
the rotor has come to a complete stop.
Loading the Rotor:
To ensure safe operation and long life of the
instrument, samples must always be loaded into
the rotor in a balanced fashion.
Tubes and strips must always be loaded
symmetrically across from one another and
contain the same volume of liquid. See below:
Examples of properly balanced
rotors:
Examples of improperly balanced
rotors:
Question and Answer IconNeed help?

Do you have a question about the Denville Scientific MiniMouse II Mini and is the answer not in the manual?

Denville Scientific MiniMouse II Mini Specifications

General IconGeneral
BrandDenville Scientific
ModelMiniMouse II Mini
CategoryLaboratory Equipment
LanguageEnglish

Summary

Rotor Exchange

Removing the Rotor from the motor shaft

The Clik-n-Lok system allows instant rotor removal by pulling upwards off the shaft; no tools are needed.

Attaching the Rotor to the motor shaft

Place rotor on the motor shaft and press down until it stops; the rotor is then secured without tools.