EasyManua.ls Logo

Desco 19787 User Manual

Desco 19787
8 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website: Desco.com
TB-3062 Page 1 of 8
© 2013 DESCO INDUSTRIES, INC.
Employee Owned
Digital Surface Resistance Meter Kit
Installation, Operation and Maintenance
Description
The Desco 19787 Digital Surface Resistance Meter Kit is
an instrument designed to measure resistance point-to-point
(Rtt) or surface to ground (Rtg) in accordance with ESD
Association Standard ANSI/ESD S4.1.
Its test funcitions include:
Resistance measurement accuracy of ±10%
(±20% for resistance values 1 x 10
11
and greater)
Resistance range of <1 x 10
3
ohms to >1 x 10
12
ohms
Under load voltages of 10 and 100 volts ±5%
Electrification period of approximately 15 seconds
The Digital Surface Resistance Meter also measures ambient
temperature and relative humidity.
The Digital Surface Resistance Meter Kit and its accessories
are available in the following item numbers:
April 2013
Figure 1. Desco Digital Surface Resistance Meter Kit
The Surface Resistance Meter Kit (or its Meter) is referenced
and designed to be used to make measurements in
accordance with the test procedures in:
Compliance Verification - ESD TR53 - Resistance
Measurements
Worksurfaces - ANSI/ESD S4.1 Worksurfaces
Floors - ANSI/ESD S7.1- Resistive Characterization of
Materials Floor Materials
Foot Grounders - ESD SP9.2 - Foot Grounders
Resistive Characterization
Garments - ANSI/ESD STM2.1 Garments
Seating - ANSI/ESD STM12.1- Seating - Resistive
Measurement
Floor/Footwear - ANSI/ESD STM97.1 - Floor Materials
and Footwear- Resistance Measurement in Combination
with a Person
“A Compliance Verification Plan shall be established
to ensure the Organization’s fulfillment of the technical
requirements of the ESD Control Program Plan. Process
monitoring (measurements) shall be conducted in
accordance with a Compliance Verification Plan that
identifies the technical requirements to be verified, the
measurement limits and the frequency at which those
verifications shall occur. The Compliance Verification Plan
shall document the test methods and equipment used for
process monitoring and measurements. If the test methods
used by the Organization differ from any of the standards
referenced in this document, then there must be a tailoring
statement that is documented as part of the ESD Control
Program Plan. Compliance verification records shall be
established and maintained to provide evidence of conformity
to the technical requirements.
The test equipment selected shall be capable of making the
measurements defined in the Compliance Verification Plan.”
(ANSI/ESD S20.20 section 7.3)
Reference Literature
In addition to those noted above:
ANSI/ESD S20.20 - Development of ESD Control Program
ESD ADV1.0 - Glossary of Terms
ANSI/ESD S6.1 Grounding
These documents can be obtained directly from the ESD
Association, 7902 Turin Rd., Suite 4, Rome, NY 13440-2069,
(315) 339-6937 or www.ESDA.org.
TECHNICAL BULLETIN TB-3062
Made in the
United States of America
Item Description
19787 Digital Surface Resistance Meter Kit
19788 Digital Surface Resistance Meter
19783 Shielded Test Leads
50003 Five Pound Electrodes, 1 Pair
50004 Replacement T-Handles for Electrodes, 1 Pair
50005 Concentric Ring Probe
Question and Answer IconNeed help?

Do you have a question about the Desco 19787 and is the answer not in the manual?

Desco 19787 Specifications

General IconGeneral
BrandDesco
Model19787
CategoryMeasuring Instruments
LanguageEnglish

Summary

Operation

General Guidelines

Provides instructions for using electrodes and general best practices for measurements.

Reporting and Using Test Results

ANSI/ESD 4.1 (Worksurfaces) Reporting

Details required reporting for resistance-to-ground and point-to-point for worksurfaces.

ANSI/ESD 7.1 (Floors) Reporting

Details required reporting for resistance-to-ground and point-to-point for floors.

Product Qualification Test Lab Test Procedure Guideline

Maintenance

Battery Replacement

Instructions for replacing the meter's AA alkaline batteries when function LEDs flash.

Calibration

Specifications

Resistance Ranges and Accuracy

Details the resistance measurement ranges, accuracy, and voltage compliance.

Related product manuals