25
E
Error Codes
If INFO appears on the screen with a Code number, perform the corresponding Corrective
Action.
Code Description Corrective Action
101 Received Signal Too Weak,
Measuring Time Too Long
Use the target plate or change the target surface.
102 Received Signal Too High Target is too reective. Use the target plate or
change the target surface.
201 Too Much Background Light Reduce the background light on the target area.
202 Laser Beam Interrupted Remove the obstacle and repeat the measurement.
301 Temperature Too High Allow the device to cool down to a temperature
within the specied Operating Temperature
Range.
302 Temperature Too Low Allow the device to warm up to a temperature within
the specied Operating Temperature Range.
401 Hardware Error Switch the device on/off several times. If the error
still occurs, return the defective device to the
Service Center or distributor. Refer to the Warranty.
402 Unknown Error Contact the Service Center or distributor. Refer to
the Warranty.
500 Data Error Contact the Service Center or distributor. Refer to
the Warranty.