EasyManua.ls Logo

dideco D902 LILLIPUT 2 ECMO User Manual

dideco D902 LILLIPUT 2 ECMO
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
LILLIPUT2 ECMO
GB – ENGLISH - INSTRUCTIONS FOR USE......................................................3
IT – ITALIANO – ISTRUZIONI PER L’USO .........................................................7
FR – FRANÇAIS - MODE D'EMPLOI.................................................................11
DE – DEUTSCH - GEBRAUCHSANWEISUNG.................................................15
NL – NEDERLANDS – GEBRUIKSAANWIJZINGEN........................................19
011548/03
Question and Answer IconNeed help?

Do you have a question about the dideco D902 LILLIPUT 2 ECMO and is the answer not in the manual?

dideco D902 LILLIPUT 2 ECMO Specifications

General IconGeneral
Branddideco
ModelD902 LILLIPUT 2 ECMO
CategoryOxygen Equipment
LanguageEnglish

Summary

English Instructions for Use

Product Description

Overview of the D 902 LILLIPUT 2 ECMO oxygenator and its integrated heat exchanger.

Technical Specifications

Details on membrane, surface area, priming volume, and connection sizes.

Intended Application

Specifies use as an infant oxygenator for ECMO and patient weight limits.

Critical Safety Information

Highlights warnings and cautions for safe device operation and potential hazards.

Label Symbols Explained

Defines symbols used on product labels for traceability and single-use indication.

Device Setup Procedures

Step-by-step guide for setting up the oxygenator holder and associated equipment.

Priming and Recirculation Steps

Instructions for priming the circuit, purging air, and performing recirculation.

Initiating Bypass Operation

Guidance on starting bypass, checking heat exchanger, and monitoring blood gases.

Management During Bypass

Procedures for low flow recirculation and managing the device during bypass.

Terminating Bypass Process

Steps for safely ending bypass, including flow adjustments and line management.

Post-Bypass Blood Recovery

Instructions for recovering blood from the circuit after bypass termination.

Oxygenator Replacement Protocol

Guidelines for oxygenator replacement based on usage duration or performance.

Compatible Medical Equipment

Lists recommended tubing, probes, and mixers for use with the device.

Manufacturer's Limited Warranty

Details the warranty terms and conditions for the medical device.