Place scale on a flat surface and press ON/OFF to power on.
Wait for the scale to complete start-up and display zero.
Position the item to be weighed on the platform.
Allow a few seconds for the display to stabilize before reading.
Press the MODE key to cycle through available units of measure.
The scale remembers the last used unit when powered off and on.
Position the empty container or vessel on the weighing platform.
Press TARE to reset the display to zero; removing it shows negative weight.
Add contents to the container; display shows net weight.
Remove items and press TARE again to return to gross weight display.
Press and hold MODE until 'CAL' appears on the display.
Press MODE again; display shows 'CAL' and required weight.
Place the correct calibration weight in the center of the platform.
Display shows 'PASS' upon successful calibration, then returns to normal.
Zero range shift. Perform calibration; persistent issue may indicate load cell damage.
Maximum capacity exceeded. Remove excess weight to prevent load cell damage.
Unstable reading. Place the scale in a more stable position.