ICS-2100 Ion Chromatography System
178 Doc. 065291-01 3/09
11. Run the pump until the mixer is filled and there is a consistent flow of
water exiting the mixer outlet.
It takes about 5 minutes to fill a 4-mm mixer and about 2 minutes to
fill a 2-mm mixer.
12. Turn off the pump.
5.22.3 Filling the EGC CO
3
Mixer with Eluent
Before initial use, follow the steps in this section to fill the EGC CO
3
Mixer with the eluent required for the application to be run. The
procedure varies slightly, depending on the type of eluent to be generated:
For carbonate eluent, see Filling the EGC CO3 Mixer with
Carbonate Eluent below.
For carbonate/bicarbonate eluent, see Filling the EGC CO3 Mixer
with Carbonate/Bicarbonate Eluent” on page 182.