Operation
Manual
Premiere 5100
Dissolution Test System

Table of Contents

Questions and Answers

  • W
    Whitney SimmonsSep 9, 2025
    What causes blurred display on Distek Test Equipment and how to fix it?
    • J
      Jean GutierrezSep 9, 2025
      To fix a blurred display on your Distek Test Equipment, adjust the display contrast by turning the multi-turn potentiometer on the stage board.
  • K
    Kimberly ThompsonAug 23, 2025
    What to do if Distek Premiere 5100 Test Equipment has no stirring?
    • C
      Cheryl RiceAug 23, 2025
      If there is no stirring, check the fuse (F5) for the motor on the motor/heater board to ensure it is not faulty.
  • V
    Veronica KimAug 25, 2025
    What to do if my Distek Premiere 5100 has no heat?
    • R
      Randy UnderwoodAug 25, 2025
      If there is no heat, check for any defective triacs that stay off all the time when heaters are enabled.
  • C
    Chloe CummingsAug 26, 2025
    What to do if Distek Premiere 5100 LED display not scrolling/locks up?
    • P
      Paul AguilarAug 27, 2025
      If the LED display is not scrolling or locks up, check the LED cable going from the motor/heater board (5103) to the LED board (5104) for any loose wires on the crimp connector.
  • C
    Connie BrooksAug 27, 2025
    What to do if Distek Test Equipment has no LED display?
    • P
      ppinedaAug 27, 2025
      If there is no LED display, check the ribbon cable that goes from the LED connector to the motor board for any loose wires or crimps.
  • J
    Jeffery PriceAug 29, 2025
    How to fix Distek Test Equipment "CRITICAL ERROR"?
    • E
      Elizabeth HessAug 29, 2025
      The "CRITICAL ERROR" indicates abnormal heating in the vessel. Ensure the calibration probe is plugged in, and then press ENTER.
  • A
    Ashley BanksSep 17, 2025
    How to level the Distek Test Equipment?
    • H
      henry60Sep 18, 2025
      To level the unit, adjust the front leveling feet under the instrument base from front to back and side to side.
  • J
    Jesse ThomasSep 18, 2025
    What causes noise in the drive assembly of Distek Test Equipment?
    • R
      ruizcollinSep 22, 2025
      If there is noise in the drive assembly, check the drive motor’s conductive brushes for wear.
  • D
    Deborah SmithSep 7, 2025
    How to troubleshoot Distek Premiere 5100 no communication (handshaking) issue?
    • A
      Angela SilvaSep 8, 2025
      If there is no communication (handshaking), check the I/O cable for any faulty connections.
  • P
    Patrick WallaceSep 4, 2025
    What to do if the shaft is not rotating but the Distek Test Equipment unit heats properly?
    • T
      Timothy DunnSep 4, 2025
      If the shaft is not rotating but the unit heats properly, check if the 0.5 amp fuse located on the 5103 PCB is blown.

Summary

Chapter 1: Introduction

Specification

Details system technical specifications like vessels, heating, RPM, display, and interfaces.

Warnings and Notes

Outlines safety warnings, high voltage risks, and operational precautions for safe use.

Chapter 2: Installation

Unpacking

Instructions for safely removing the instrument from its packaging and checking for damage.

Leveling: Model 5100

Steps to ensure the instrument is level on the bench for accurate operation.

Paddles and Baskets

Procedures for installing and handling paddle blades and baskets for dissolution testing.

Installation Checkout

Critical operational parameter checks after installation to detect shipping or setup issues.

Chapter 3: Operation

Control Keypad

Explanation of the functions of each key on the instrument's control panel.

Display and Control Screens

Overview of the instrument's LCD display and various operational screens.

Automatic Testing Methods

Guidance on setting up and running pre-programmed dissolution test methods.

Manual Operation

Procedures for controlling instrument parameters like RPM and temperature manually.

System Setup and Calibration

Essential procedures for calibrating temperature probes for accurate measurements.

Chapter 4: Maintenance

General Care and Maintenance of Model 5100

Guidelines for routine cleaning, lubrication, and care of the instrument.

Chapter 5: Troubleshooting & Service

Troubleshooting Chart

A guide to identifying and resolving common instrument problems and their solutions.

Model 5100 System Initialize Procedure

Steps to initialize the instrument, often after chip replacement or heating issues.

Service

Contact information and procedures for obtaining technical support or arranging repairs.

Appendix A: Spare Parts and Accessories

Baskets

List of available baskets with different mesh sizes and coatings.

Kits & Accessories

Various kits and accessories offered, including press, visual aids, and manuals.

Service Parts

List of replacement parts for the instrument, including boards, fuses, and belts.

Validation Tools

Tools for instrument validation and calibration, such as tachometers and wobble meters.

Appendix F: Sample Printouts from Parallel Printer

Method Printout

Example of a printout showing test method settings and parameters.

Test Summary Printout

Example of a printout summarizing test results, including RPM and temperature data.

InstaPrint

Example of an InstaPrint output showing real-time temperature and RPM data.

Distek Premiere 5100 Specifications

General IconGeneral
ModelPremiere 5100
TypeDissolution Test System
ManufacturerDistek
Number of Stations6 or 8
ComplianceUSP, EP, JP
Stirring Speed25 to 250 RPM
Basket Speed Range25 to 250 RPM
Paddle Speed Range25 to 250 RPM
Temperature Accuracy±0.1°C
Temperature Resolution0.1°C
Vessel Volume1000 mL
Vessel MaterialBorosilicate Glass
Power Requirements100-240 VAC, 50/60 Hz
DisplayTouchscreen
Data OutputUSB, Ethernet
Vessels6 or 8